Integrated Wellbore Caliper
Systems and methods are provided for an integrated caliper measurement obtained based on caliper measurements from at least two downhole tools. In one example, a system for obtaining such an integrated caliper includes several caliper tools and data processing circuitry. The caliper tools may respectively obtain caliper measurements of a wellbore. The data processing circuitry may assign respective weights to the respective caliper measurements and combine the weighted caliper measurements to obtain an integrated caliper measurement. The confidence factors may be determined based at least partly on at least two of a plurality of weighting functions. These weighting functions may each relate a confidence of the caliper measurements to an environmental or drilling characteristic occurring when the caliber measurements are obtained.
This disclosure relates to an integrated caliper measurement of a wellbore obtained using weighted values of different caliper measurements of the wellbore.
This section is intended to introduce the reader to various aspects of art that may be related to various aspects of the present techniques, which are described and/or claimed below. This discussion is believed to be helpful in providing the reader with background information to facilitate a better understanding of the various aspects of the present disclosure. Accordingly, it should be understood that these statements are to be read in this light, and not as admissions.
Logging-while-drilling (LWD) tools are downhole tools that can obtain logging measurements while a wellbore is being drilled. LWD tool may collect information about the geological formation surrounding the wellbore as well as information relating to the wellbore itself. One such LWD tool is known as an LWD caliper tool, which provides a measurement of the diameter of the wellbore at various depths. Caliper measurements are used in a variety of operations, including drilling, cementing, and evaluation of the geological formation. During drilling, for example, the caliper data can be used to monitor the wellbore condition (e.g., identifying possible wellbore washout and/or impending wellbore instability), thus allowing the driller to take remedial action. During well completion, the caliper data can be used to accurately evaluate the volume of cement to fill the casing annulus, as well as aiding in the selection of casing points. A reliable caliper may also be useful during logging to correct some formation evaluation measurements for wellbore size and to evaluate the quality of almost all other LWD logs. The caliper measurements can also be used to plan services, such as dip meters or formation testers, that may be affected by unfavorable wellbore conditions.
A variety of tools have been developed to obtain LWD caliper measurements, including ultrasonic tools and radiation-based density tools. None of these tools, however, can provide a reliable answer in all drilling environments and conditions. For example, a caliper measurement from the ultrasonic tool has a range that is limited by eccentering, mud weight, and acoustic impedance contrast between drilling mud and the geological formation. Meanwhile, the radiation-based caliper measurements may be available substantially only when the radiation-based tool is rotating. Thus, in the often-valuable sections of the geological formation where the LWD tool is “sliding”—moving through the geological formation without having to rotate—the radiation-based density caliper measurement may not be valid. While calipers from different types of tools may be accurate at different times and under different conditions, any single caliper from one particular tool may not be trustworthy under all wellbore and operating conditions.
SUMMARYA summary of certain embodiments disclosed herein is set forth below. It should be understood that these aspects are presented merely to provide the reader with a brief summary of these certain embodiments and that these aspects are not intended to limit the scope of this disclosure. Indeed, this disclosure may encompass a variety of aspects that may not be set forth below.
Embodiments of the disclosure relate to systems and methods for an integrated caliper measurement obtained based on caliper measurements from at least two downhole tools or techniques. In one example, a system for obtaining such an integrated caliper includes several caliper tools and data processing circuitry. The caliper tools may respectively obtain caliper measurements of a wellbore. The data processing circuitry may assign respective weights to the respective caliper measurements and combine the weighted caliper measurements to obtain an integrated caliper measurement.
In a second example, one or more tangible non-transitory machine-readable media may store instructions to receive a first caliper measurement deriving from a first downhole tool, a second caliper measurement deriving from a second downhole tool, and one or more conditions under which the first downhole tool obtained the first caliper measurement and under which the second downhole tool obtained the second caliper measurement. The media may also store instructions to determine a first confidence factor—representing a degree of confidence that the first caliper measurement is accurate based at least in part on the one or more conditions—and a second confidence factor—representing a degree of confidence that the second caliper measurement is accurate based at least in part on the one or more condition. The media may also store instructions to determine an integrated caliper measurement using the first caliper measurement, the first confidence factor, the second caliper measurement, and the second confidence factor.
In a third example, a method includes obtaining a first caliper measurement from a first downhole tool at a depth in a wellbore and obtaining a second caliper measurement from a second downhole tool at the depth in the wellbore. First and second weight factors may be determined. The first weight factors are associated with a reliability of the first caliper measurement under the conditions in which the first caliper measurement was obtained. The second weight factors are associated with a reliability of the second caliper measurement under conditions in which the second caliper measurement was obtained. First and second confidence factors may be determined. The first confidence factor may be determined based at least partly on the first plurality of weight factors, and the second confidence factor may be determined based at least partly on the second plurality of weight factors. Based at least partly on the first caliper measurement, the second caliper measurement, the first confidence factor, and the second confidence factor, an integrated caliper measurement may be determined.
Various refinements of the features noted above may exist in relation to various aspects of this disclosure. Further features may also be incorporated in these various aspects as well. These refinements and additional features may exist individually or in any combination. For instance, various features discussed below in relation to one or more of the illustrated embodiments may be incorporated into any of the above-described aspects of this disclosure alone or in any combination. The brief summary presented above is intended to familiarize the reader with certain aspects and contexts of embodiments of this disclosure without limitation to the claimed subject matter.
Various aspects of this disclosure may be better understood upon reading the following detailed description and upon reference to the drawings in which:
One or more specific embodiments of the present disclosure will be described below. These described embodiments are examples of the presently disclosed techniques. Additionally, in an effort to provide a concise description of these embodiments, certain features of an actual implementation may not be described in the specification. It should be appreciated that in the development of any such actual implementation, as in any engineering or design project, numerous implementation-specific decisions may be made to achieve the developers' specific goals, such as compliance with system-related and business-related constraints, which may vary from one implementation to another. Moreover, it may be appreciated that such a development effort might be complex and time consuming, but would nevertheless be a routine undertaking of design, fabrication, and manufacture for those of ordinary skill having the benefit of this disclosure.
When introducing elements of various embodiments of the present disclosure, the articles “a,” “an,” and “the” are intended to mean that there are one or more of the elements. The terms “comprising,” “including,” and “having” are intended to be inclusive and mean that there may be additional elements other than the listed elements. Additionally, it should be understood that references to “one embodiment” or “an embodiment” of the present disclosure are not intended to be interpreted as excluding the existence of additional embodiments that also incorporate the recited features.
This disclosure relates to an integrated caliper measurement obtained by combining caliper measurements from two or more downhole tools. Although one caliper measurement from a single tool may not always be accurate under certain conditions, the integrated caliper measurement of this disclosure represents a combined caliper measurement that generally may be better than any one caliper measurement taken individually. Specifically, at least two caliper measurements derived from two different tools or techniques may be obtained. Depending on the conditions under which the caliper measurements were obtained, the individual caliper measurements may be weighted according to the likelihood—under these conditions—that the caliper measurement is reliable. For instance, a number of weight factors may be assigned to each caliper based on the reliability of the caliper measurement under respective conditions. The product of these weight factors may provide total confidence factors associated with the overall reliability of each caliper measurement under those conditions. Based on the confidence factors and the caliper measurements, the integrated caliper measurement may be obtained that outperforms the reliability of either of the composite caliper measurements individually.
Such an integrated caliper measurement system may be used in a logging-while-drilling (LWD) setting.
The environment of the wellbore 14 may vary widely depending upon the location and situation of the geological formation 12. For example, rather than a land-based operation, the wellbore 14 may be drilled into the geological formation 12 under water of various depths, in which case the surface 16 may include topside equipment such as an anchored or floating platform, and some of the components used may be positioned at or near a point where the wellbore 14 enters the earth beneath a body of water. Moreover, in the example of
As illustrated in
The BHA 34 may include more or fewer LWD tools 40 than those shown in
One example of a logging-while-drilling (LWD) integrated caliper tool 40 appears in
Data processing circuitry 54 may be located within the LWD integrated caliper tool 40 or elsewhere (e.g., at the surface 16). The data processing circuitry 54 may include a microprocessor (μp) 56, memory (M) 58, and/or storage (S) 60. The data processing circuitry 54 may obtain caliper measurements from the first caliper tool 50 and the second caliper tool 52. Weighting these measurements based on the conditions in the wellbore 14, the data processing circuitry 54 may provide an integrated caliper answer 62. The integrated caliper answer 62 may generally be more accurate over all ranges of depths in the wellbore 14 than the first caliper tool 50 or the second caliper tool 52 alone. The microprocessor 56 may determine the integrated caliper answer 62 based on instructions stored in the memory 58 or the storage 60. The memory 58 and/or the storage 60 thus may represent any suitable article of manufacture that can store these instructions, such as read-only memory (ROM), random access memory (RAM), flash memory, optical storage media, or a hard disk drive, to name a few examples.
The LWD integrated caliper tool 40 may include an ultrasonic caliper tool 70 and a gamma density tool 72, from which to obtain two caliper measurements, in the example of
The gamma density tool 72 may obtain a caliper measurement derived from a gamma density measurement. The gamma density measurement may be determined from the Compton scattering of gamma-rays that are emitted by a gamma-ray source 76. The gamma-rays emitted by the gamma-ray source 76 may exit the gamma density tool 72 to interact with the materials surrounding the LWD integrated caliper tool 40. Some of the gamma-rays emitted by the gamma-ray source 76 may be detected by a short spaced (SS) detector 78 and others may be detected by a long spaced (LS) detector 80. Based on the number of gamma-rays detected at the SS detector 78 and the LS detector 80, a density measurement may be derived, from which a caliper measurement of the wellbore 14 can be determined. The ultrasonic caliper tool 70 and the gamma density tool 72 may obtain caliper measurements in any suitable fashion, including those well known in the art.
The caliper measurements obtained by the ultrasonic caliper tool 70 and the gamma density tool 72 may respectively prove most reliable under different circumstances. For example, the caliper measurements made by the ultrasonic caliper tool 70 are direct measurements that are azimuthal and may be available in real time. The caliper measurements from the ultrasonic caliper tool 70 may work in drilling fluid 24 of both oil-based mud (OBM) and water-based mud (WBM) and may have a precision of between 0.1 and 0.2 inches in some cases. There are adverse factors, however, that may affect the caliper measurements from the ultrasonic caliper tool 70. These adverse factors include, among other things, the acoustic impedance contrast between the geological formation 12 and the drilling fluid 24, the attenuation of the acoustic signal on which the caliper measurement is based when the drilling fluid 24 is heavy mud, the impact of rugosity of the wellbore 14 on the caliper measurement, and the alignment of the transducers 74 to the wall 46 of the wellbore 14.
For instance, one limiting factor for the caliper measurement from the ultrasonic tool 70 is the mud weight of the drilling fluid 24. Heavy mud attenuates the sonic pulse of the ultrasonic tool 70, reducing the amount of measurable standoff. Also, the strength of the reflected sonic wave output by the ultrasonic tool 70 decreases as the acoustic impedance of the drilling fluid 24 approaches that of the geological formation 12, often preventing precise echo detection, and thus a precise caliper measurement. Other effects that compromise the accuracy of the caliper from the ultrasonic caliper tool 70 include eccentering (e.g., as caused by misalignment between the transducers 74 and the wellbore 14); the impact of cuttings and the rugosity of the wall 46 of the wellbore 14, both of which scatter and reflect the sonic wave of the ultrasonic caliper tool 70; and the effect of gas traveling through the wellbore 14. Moreover, when the ultrasonic caliper tool 70 includes only a single transducer 74, the ultrasonic caliper tool 70 may not get meaningful results when the LWD integrated caliper tool 40 is not rotating. Finally, the accuracy of the ultrasonic caliper tool 70 may depend on the value assumed of the speed of sound through the drilling fluid 24 as used to calculate the caliper measurement from the ultrasonic caliper tool 70. The speed of sound in the drilling fluid 24 may be determined as a function of several environmental parameters, such as mud type, mud density, mud salinity, pressure, temperature, and gas cut. These parameters may be derived from different sources, however, and they may not represent the actual downhole conditions in the wellbore 14. Furthermore, no complete database of sound speed in various drilling fluids 24 is believed to be available, and thus this parameter may be inferred by extrapolation. When the speed of sound is inadvertently wrongly assumed, the caliper measurement from the ultrasonic caliper tool 70 may be inaccurate.
The caliper measurement from the gamma density tool 72 may similarly have strengths and weaknesses. Considering strengths, the caliper derived from the gamma density tool 72 may provide azimuthal measurements and may work when the drilling fluid 24 is oil-based mud (OBM) or water-based mud (WBM). The caliper measurement from the gamma density tool 72 also has good sensitivity to standoff and works in the presence of gas, where the caliper measurement from the ultrasonic tool 70 may sometimes fail. The caliper measurement from the gamma density tool 72 may be limited, however, by certain environmental conditions. For example, a high-barite-weight drilling fluid 24 may cause the caliper measurements from the gamma density tool 72 to be inaccurate. Moreover, the gamma density tool 72 may obtain a good caliper measurement primarily when the stabilizer 44 of the drill string 18 is close in size to the drill bit 20. Furthermore, the theoretical limit of the caliper measurement from the gamma density tool 72 may correspond to the depth of investigation of the LS detector 80. In one example, the results may be reliable up to a maximum of 3 inches under some conditions.
Finally, the caliper measurements from the gamma density tool 72 may be azimuthally available only when the LWD integrated caliper tool 40 is rotating. As such, the caliper measurement from the gamma density tool 72 may not be accurate when the drill string 18 is sliding. As noted above, the term “sliding” refers to the non-rotation of the bottomhole assembly (BHA) 34. This may occur when drilling with a mud motor, tripping into a well, or tripping out of a well, and often happens while traversing valuable depths of the wellbore 14. The accuracy of the caliper from the gamma density tool 72 may also depend on the quality of the measurement used as representative of the density of the geological formation 12, which itself may depend on the accuracy of other downhole tools.
To summarize, calipers from different tools, such as the ultrasonic caliper tool 70 and the gamma density tool 72, each may provide reliable measurements under certain conditions. Some of these conditions may overlap, as generally illustrated in
Similarly, a plot 110 of
These and other limiting factors that impair the caliper measurements from the ultrasonic caliper tool 70 and the caliper from the gamma density tool 72 are generally shown in the following table:
In Table 1 above, several limiting factors are listed that respectively may affect the accuracy of calipers from the ultrasonic caliper tool 70 or from the gamma density tool 72. An “X” indicates that the listed limiting factor affects caliper measurements from the ultrasonic caliper tool 70 or the gamma density tool 72. As mentioned above, these limiting factors include the contrast between the drilling fluid 24 and the geological formation 12 (mud/formation contrast), the depth of investigation (DOI) of the respective tools 70 and 72, the photoelectric (Pe) factor associated with the drilling fluid 24, the weight of the drilling fluid 24 (mud weight), the salinity of the drilling fluid 24 (mud salinity), the rugosity of the wellbore 14, the amount of eccentering of the LWD integrated caliper tool 40, and/or whether or not the BHA 34 is sliding.
As indicated by
All of the above-described scenarios are, of course, simplified to some degree. Reality is more complex, and several other factors may be considered. The full spectrum of measurements acquired by the tools of the BHA 34 and the knowledge of the environmental parameters may permit estimation of the quality of the caliper measurements from each respective tool in many different circumstances. Because different caliper measurements from different caliper tools may be more or less appropriate under different circumstances, the LWD integrated caliper tool 40 may combine the caliper measurements derived from these various techniques (e.g., from the ultrasonic caliper tool 70 and the gamma density tool 72) that may, in combination, have an enhanced accuracy and wider applicability range compared to the caliper measurements taken individually.
Indeed, as seen by a flowchart 130 of
In this way, the integrated caliper answer 62 may be thought of as generally the best available caliper from those collected by the LWD integrated caliper tool 40 at a particular depth, depending on the conditions under which the caliper measurements are obtained. The contribution of each caliper measurement in the ultimate integrated caliper answer 62 may depend on the confidence factor associated with each caliper measurement as determined at block 136. As will be discussed below, there may be one confidence factor per caliper measurement, each confidence factor being determined based on one or a group of weight factors.
The weight factors used to ascertain the confidence factors may be values between 0 and 1 that qualify the conditions in which the LWD integrated caliper tool 40 is working. These values may be obtained from characteristics curves that represent the effect of a particular parameter on the particular caliper measurement (e.g., the effect of tool sliding on the reliability of the caliper from the gamma density tool 72). In one example, these curves or functions may take the following form:
where the terms a and b are constants that control the slope of the curve of the weight factor, and x is the parameter to evaluate.
Every weight factor may have a particular function shape associated with it. Namely, each weight factor may be computed as a function of a particular parameter or measurement that defines the function shape. The functions may be derived from experimental databases, measurement specifications, field data, and/or computer modeling. Some examples of functional forms of weight factors appear in
In
An example of how the weight factors may be used to obtain the integrated caliper answer 62 appears in a system 170 of
In the particular example shown in
The second inputs 174 may include the standoffs of the gamma density tool 72, the collar rotation speed (CRPM), the rate of penetration of the drill string 18 into the wellbore 14 (ROP), a size of a stabilizer of the drill string 18, a size of the stabilizer 44 of the drill string 18, a mud weight of the drilling fluid 24, a mud type of the drilling fluid 24 (e.g., oil-based mud (OBM) or water-based mud (WBM)), a size of the drill bit 20, and/or a slowness of the drilling fluid 24. In addition, the second inputs 174 may include a bottom bulk density (ROBB), a bottom bulk density correction (DRHB), and a volumetric photoelectric factor value (U), which may be obtained by the gamma density tool 72 or any other suitable downhole tool. These additional inputs may be used to correct the caliper measurement from the gamma density tool 72 in addition to determining the second weighting values 178. The second inputs 174 may also include an average caliper measurement value from the gamma density tool 72 (DCAV). The second inputs 174 may be used to determine the second weight factors 178.
The weight factors 176 and 178 relate the reliability of the ultrasonic caliper tool 70 and gamma density tool 72, respectively, depending on the conditions indicated by the first inputs 172 and the second inputs 174, respectively. Examples of some weight factors 176 and 178 are provided below in Table 2:
In Table 2 above, the first weight factors 176 are ultrasonic caliper weight factors denoted by “US” and the second weight factors 178 are density-derived caliper weight factors denoted by “DEN.” In one embodiment, the weight factors 176 and 178 illustrated in
Separate confidence factors relating to each caliper measurement may be ascertained as the product of all of the individual weight factors 176 or 178 relating to that caliper measurement. In the example of
The values nucqf and ndcqf may be referred to as normalized confidence factors associated with the ultrasonic caliper tool 70 and the gamma density tool 72, respectively. The sum of the normalized confidence factors nucqf and ndcqf is 1. In other words, the higher the confidence factor associated with a caliper from a particular tool (e.g., the ultrasonic caliper tool 70 or the gamma density tool 72), the greater the contribution of that caliper measurement to the final value of the integrated caliper answer 62. The integrated caliper answer 62 appears in
ICAL=ndcqf·DCAV+nucqf·UCAV.
The resulting integrated caliper answer 62 (ICAL) may represent the most appropriate measurement from among the various caliper measurements available through the data set being analyzed. The value of the integrated caliper answer 62 from the LWD integrated caliper tool 40 has been born out through several case studies, which will be described below. In a first case study, a data set including caliper measurements from an ultrasonic caliper tool 70 and caliper measurements from a gamma density tool 72 were considered in a logging-while-drilling setting with the following conditions:
In the earlier areas of the depth interval shown in the plot 180 of
The accuracy of the respective caliper measurements is reflected in the confidence factors plotted in plot 191. In the depths where the average density caliper (DCAV) of curve 186 appears (inaccurately) to indicate a washout, the density caliper confidence factor (DCQF) of curve 195 is at 0. Even so, the ultrasonic caliper confidence factor (UCQF) of curve 196 and integrated caliper confidence factor (ICQF) of curve 197 remain high.
As can be seen in
A second case study also illustrates the effectiveness of the integrated caliper technique of this disclosure. The second case study involves conditions as provided by Table 4 below:
An example of results obtained in this second case study appear in
As seen in
Although the integrated caliper measurement of this disclosure has been described in a logging-while-drilling (LWD) implementation, the integrated caliper system of this disclosure may be used with any suitable means of conveyance and should not be understood to be limited as such. Moreover, while the integrated caliper measurement of this disclosure may benefit from component caliper measurements obtained by different tools in the same sub and/or logging run, caliper data from different tools and/or logging runs may be used in a post-hoc determination of the integrated caliper measurement at a later time (provided the caliper measurements by multiple tools can be aligned by depth and the particular environmental conditions associated with each caliper measurement can be ascertained or estimated).
The specific embodiments described above have been shown by way of example, and it should be understood that these embodiments may be susceptible to various modifications and alternative forms. Finally, it should be further understood that the claims are not intended to be limited to the particular forms disclosed, but rather to cover modifications, equivalents, and alternatives falling within the spirit and scope of this disclosure.
Claims
1. A system comprising:
- a plurality of caliper tools configured to obtain respective caliper measurements of a wellbore; and
- data processing circuitry configured to assign respective confidence factors to weight the respective caliper measurements and combine the weighted caliper measurements to obtain an integrated caliper measurement, wherein the confidence factors are determined based at least partly on at least two of a plurality of weighting functions that each relate a confidence of the caliper measurements to an environmental or drilling characteristic occurring when the caliber measurements are obtained.
2. The system of claim 1, wherein the plurality of caliper tools comprises an ultrasonic caliper tool configured to obtain an ultrasonic caliper.
3. The system of claim 1, wherein the plurality of caliper tools comprises a gamma density tool configured to obtain a density-derived caliper.
4. The system of claim 1, wherein the plurality of caliper tools comprises three or more caliper tools that provide three or more respective caliper measurements.
5. The system of claim 1, wherein the data processing circuitry is configured to assign the respective confidence factors to the respective caliper measurements based at least in part on a characteristic of drilling fluid in the wellbore.
6. The system of claim 5, wherein the characteristic of the drilling fluid in the wellbore comprises a mud weight of the drilling fluid, a mud type of the drilling fluid, a mud slowness of the drilling fluid, or a photoelectric factor associated with the drilling fluid, or any combination thereof.
7. The system of claim 1, wherein the data processing circuitry is configured to assign the respective confidence factors to the respective caliper measurements based at least in part on a characteristic of operation of a drill string that houses the plurality of caliper tools.
8. The system of claim 7, wherein the characteristic of operation of the drill string comprises a rotation speed of the drill string, an amount of eccentering of the drill string in the wellbore, or a rate of penetration of the drill string in the wellbore, or any combination thereof.
9. The system of claim 1, wherein the data processing circuitry is configured to assign the respective confidence factors to the respective caliper measurements based at least in part on a condition of the wellbore or of a drill string that houses the plurality of caliper tools, wherein the condition is configured to be input by a human operator of the system.
10. The system of claim 1, comprising at least one other tool configured to ascertain a condition of the wellbore or of a drill string that houses the plurality of caliper tools, wherein the data processing circuitry is configured to assign the respective weights to the respective caliper measurements based at least in part on the condition of the wellbore.
11. One or more tangible non-transitory machine-readable media storing instructions to:
- receive a first caliper measurement deriving from a first downhole tool;
- receive a second caliper measurement deriving from a second downhole tool;
- receive a plurality of conditions under which the first downhole tool obtained the first caliper measurement and under which the second downhole tool obtained the second caliper measurement;
- determine a first confidence factor representing a degree of confidence that the first caliper measurement is accurate based at least in part on a plurality of weighting factors each associated with one of the plurality of conditions;
- determine a second confidence factor representing a degree of confidence that the second caliper measurement is accurate based at least in part on a plurality of weighting factors each associated with one of the plurality of conditions; and
- determine an integrated caliper measurement using the first caliper measurement, the first confidence factor, the second caliper measurement, and the second confidence factor.
12. The machine-readable media of claim 11, wherein:
- the instructions to determine the first confidence factor comprise instructions to determine a first plurality of weight factors each relating to a response of the first downhole tool to at least two of the plurality of conditions and determine the first confidence factor based at least in part on the first plurality of weight factors; and
- the instructions to determine the second confidence factor comprise instructions to determine a second plurality of weight factors each relating to a response of the second downhole tool to at least two of the plurality of conditions and determine the second confidence factor based at least in part on the second plurality of weight factors.
13. The machine-readable media of claim 12, wherein:
- the instructions to determine the first confidence factor comprise instructions to multiply the first plurality of weight factors together, wherein each of the first plurality of weight factors represents a value between 0 and 1; and
- the instructions to determine the second confidence factor comprise instructions to multiply the second plurality of weight factors together, wherein each of the second plurality of weight factors represents a value between 0 and 1.
14. The machine-readable media of claim 11, wherein the instructions to determine the integrated caliper measurement comprise instructions to weight the first caliper measurement based on a relationship between the first confidence factor to the second confidence factor and to weight the second caliper measurement based on a relationship between the second confidence factor to the first confidence factor.
15. The machine-readable media of claim 11, wherein the instructions to determine the integrated caliper measurement accord with the following relationships: I C A L = ndcqf · D C A V + nucqf · U C A V; ndcqf = D C Q F D C Q F + U C Q F; and nucqf = U C Q F D C Q F + U C Q F;
- where ICAL represents the integrated caliper measurement, DCAV represents the first caliper measurement, UCAV represents the second caliper measurement, DCQF represents the first confidence factor, and UCQF represents the second confidence factor.
16. The machine-readable media of claim 11, comprising instructions to determine a confidence factor of the integrated caliper measurement in accordance with the following relationships: I C Q F = ndcqf · D C Q F + nucqf · U C Q F; ndcqf = D C Q F D C Q F + U C Q F; and nucqf = U C Q F D C Q F + U C Q F;
- where ICQF represents the confidence factor of the integrated caliper measurement, DCQF represents the first confidence factor, and UCQF represents the second confidence factor.
Type: Application
Filed: Dec 17, 2013
Publication Date: Nov 5, 2015
Inventor: Mauro Manclossi (Paris)
Application Number: 14/651,664