Dispersive near-infrared spectrometer with automatic wavelength calibration
The present invention is a dispersive, diffraction grating, NIR spectrometer that automatically calibrates the wavelength scale of the instrument without the need for external wavelength calibration materials. The invention results from the novel combination of: 1) a low power He—Ne laser at right angles to the source beam of the spectrometer; 2) a folding mirror to redirect the collimated laser beam so that it is parallel to the source beam; 3) the tendency of diffraction gratings to produce overlapping spectra of higher orders; 4) a “polka dot” beam splitter to redirect the majority of the laser beam toward the reference detector; 5) PbS detectors and 6) a software routine written in Lab VIEW that automatically corrects the wavelength scale of the instrument from the positions of the 632.8 nm laser line in the spectrum.
This application is based on and claims priority from U.S. patent application Ser. No. 10/890,942 filed Jul. 14, 2004, which is a continuation of U.S. patent application Ser. No. 10/093,584 filed Mar. 8, 2002, which claims priority to Provisional patent Application Ser. No. 60/274,300 filed on Mar. 8, 2001.
BACKGROUND OF THE INVENTIONThe present invention relates to a Near-infrared (NIR) spectrometer with automatic wavelength calibration without the need of external calibrating. NIR spectroscopy is the measurement of the wavelength and intensity of the absorption of near-infrared light by a sample. Near-infrared light spans the 800 nm-2.5 micrometers (μm) range and is energetic enough to excite overtones and combinations of molecular vibrations to higher energy levels. NIR spectroscopy is typically used for quantitative measurement of organic functional groups, especially O—H, N—H, and C—H. Analyte detection limits are typically 0.11%.
NIR spectroscopy has been shown to be a powerful analytical tool for the analysis of agricultural products, food products, petroleum products, and pharmaceuticals products. Recently, NIR spectroscopy has been approved for the analysis of pharmaceutical products, a factor that is likely to dramatically extend the number of applications of the technique. In general, when NIR spectroscopy is combined with multivariate calibration procedures, the analytical methodology that results is rapid, accurate, and requires virtually no sample preparation.1
In conventional NIR spectroscopy, a multivariate statistical model is developed that attempts to correlate subtle changes in the NIR spectrum with known compositional changes determined by standard analytical technology. Once a robust model has been developed, NIR spectroscopic measurements can be substituted for the more time consuming, labor-intensive conventional analytical measurements.2 To be completely useful, however, a model developed on one spectrometer in the laboratory should be capable of being used on different spectrometers without having to go through the model development all over again with the new instrument. To transfer a model from one spectrometer to another successfully, both instruments must ideally be identical.3
Many NIR spectrometers in use today employ dispersive systems that use diffraction grating monochromators. For these instruments, accurate wavelength calibration is important if the calibration models developed in the laboratory are to be used successfully on other instruments in the production environment. If the wavelength scales of different spectrometers are miscalibrated (as they inevitably are), problems with calibration transfer will occur.4 Because of this, the standardization of NIR spectrometers has been pursued. The rational behind this being that if instruments are alike and remain stable enough, calibration transfer no longer becomes an analytical performance issue. Instrument standardization helps ensure that spectra produced from different instruments of the same design are essentially identical. In order to successfully carry out the various instrument standardization protocols, such as those suggested by Workman and Coates5 and Wang, et al.6, it is necessary to develop strategies that would accurately characterize all the instrumental variables of importance (i.e., wavelength and photometric accuracy, spectral bandwidth, and stray light). One way to avoid this problem is to use a wavelength standard to validate the wavelength scale of the spectrometer. Various wavelength standards exist.7-10
Recently, Busch and co-workers have proposed the use of trichloromethane as a substance with sharper, isolated absorption bands that are suitable for wavelength calibration of spectrometers in the NIR region.11 The study of the use of trichloromethane as a wavelength standard showed that calibration of the wavelength scale of NIR instruments is absolutely essential, and a typical dispersive NIR spectrometer may be off by as much as 12 nm in the NIR region. Busch and co-workers have also assembled a research-grade NIR spectrometer that has been designed to allow the effect of various instrumental parameters on spectrometer performance to be studied in a systematic fashion. This is the same NIR spectrometer used to study the role of trichloromethane as a wavelength standard for NIR spectroscopy and to evaluate the stray light level in dispersive NIR spectrometers that has been designed to allow the effect of various instrumental parameters on spectrometer performance to be studied in a systematic fashion.12 This disclosure describes a novel, dispersive, diffraction grating, NIR spectrometer that automatically calibrates the wavelength scale of the instrument without the need for external wavelength calibration materials.
SUMMARY OF THE INVENTIONIn accordance with the above and related objects, the present invention is a dispersive, diffraction grating, NIR spectrometer that automatically calibrates the wavelength scale of the instrument without the need for external wavelength calibration materials. In a preferred embodiment, the present invention results from the novel combination of: 1) a low power He—Ne laser at right angles to the source beam of the spectrometer (
The present invention relates to a novel dispersive NIR spectrometer that automatically calibrates the wavelength scale of the instrument without the need for external wavelength calibration materials.13 This NIR spectrometer with automatic calibration as disclosed herein was developed by the inventor in the laboratory from commercially available component parts in novel combination. The spectrometer is in part based on a spectrometer that has been described previously.14 The general layout of a spectrometer 10 is seen in
A specially constructed sample compartment 16,
A preferred embodiment of the present invention is also shown in
The basic concept behind the laser wavelength calibration system described here is to use a 0.5-mW He—Ne laser 40,
The success of the laser wavelength calibration system derives from a combination of factors. First, radiation from a laser 40 is used to provide a small-diameter, highly collimated beam of radiation at an accurately known wavelength, for example, 632.8 nm. Radiation from a He—Ne laser 40 is reflected orthogonally by a small folding mirror 48 so that the laser radiation is co-linear with the beam from the primary QTH lamp 1. The small folding mirror 48 is small (˜3 mm diameter) so that it blocks only a tiny fraction of the primary source beam from the QTH lamp 1. Both beams are modulated simultaneously by the rotary chopper 42 and enter the monochromator 6 equipped with the diffraction grating 8.
According to the normal diffraction grating equation 1, mλ=d(sin i±sin θ), where m is the diffraction order, λ, is the wavelength, d is the grating constant, i is the angle of incidence, and, θ, is the angle of diffraction. According to Eqn. 1, for a given diffraction grating with fixed i and θ, m1λ1=m2 λ2. This means that 632.8 nm radiation in the second order will appear at the same position as 1266 nm radiation in the first order. Table I gives the apparent positions of 632.8 nm radiation for spectral orders out to six.
It is clear from Table I that the apparent locations of the 632.8 nm laser line are integral multiples of 632.8 and are, therefore, spread uniformly throughout the spectrum at m(632.8) nm, where m is the diffraction order in Eqn. 1. Because the optics of diffraction gratings are well known, the positions of the various spikes can be predicted with great accuracy.
In laboratory study, the so-called “polka-dot” beam splitter 20,
For first-order wavelengths that coincide with the higher diffraction order positions of the 632.8 nm laser line as given in Table I, the intensity of radiation striking the PbS reference detector 24 will go up (i.e., it will consist of radiation from both the QTH lamp 1 and the laser 40). Since absorbance is defined as log (Ireference/Isample), an increase in Ireference will produce an apparent increase in the absorbance at the wavelengths given in Table I. This will result in absorbance spikes at positions given by m(632.8 nm) in the spectrum, where m is an integer.
Table II lists some absorption bands of chloroform recorded with the modified NIR spectrometer that incorporates the He—Ne wavelength marker system and gives the wavelength reproducibility of the prototype instrument.
Unlike the instant invention, calibration of the wavelength scale of a FT-NIR spectrometer is often necessary due to small, inevitable misalignments of the He—Ne reference laser which introduce small wavelength shifts in the interferogram that compromise the wave number accuracy of the FT-NIR spectrometer23. Tests with the invention assembled in the laboratory have revealed that the laser wavelength calibration system performs comparably to a FT-NIR spectrometer when used to determine the absorption wavelengths for trichloromethane. Table III compares the wavelength accuracy of the laser spectrometer with a commercial Fourier transform NIR spectrometer.
While the invention has been described with a certain degree of particularity, it is manifest that many changes may be made in the arrangement of components without departing from the spirit and scope of this disclosure. It is understood that the invention is not limited to the embodiments set forth herein for purposes of exemplification, but is to be limited only by the scope of the attached claim or claims, including the full range of equivalency to which each element thereof is entitled.
Design and Evaluation of a Near-Infrared Dispersive Spectrometer that Uses a He—Ne Laser for Automatic Internal Wavelength CalibrationA diffraction-grating near-infrared spectrometer that uses a He—Ne laser for automatic internal wavelength calibration is described. The instrument uses the known location of the higher diffraction orders of the 632.8 nm laser line to perform wavelength calibration in the near-infrared region with a program written in LabVIEW. The wavelength accuracy of the dispersive spectrometer was compared with that of a Fourier-transform near-infrared spectrometer whose wavelength scale was validated by calibration with the known spectrum of ethyne. The average absolute wavelength deviation between the two spectrometers for four isolated bands of trichloromethane was found to be ±0.12 nm. The average values of the wavelengths of four isolated bands of trichloromethane obtained with the two spectrometers used in this study were determined to be: 1151.62±0.28 nm (3ν1), 1410.74±0.52 nm (2ν1+ν4), 1692.38±0.49 nm (2ν1), and 1860.20±0.16 nm (ν1+2ν4)
Index Headings Near Infrared Spectroscopy; Wavelength Calibration; Spectrometer calibration; NIR spectrum of trichloromethane.
INTRODUCTIONNear-infrared (NIR) spectroscopy has been shown to be a powerful analytical tool for the analysis of agricultural products, food products, petroleum products, and pharmaceuticals1,2. In general, when NIR spectroscopy is combined with multivariate calibration procedures, the analytical methodology that results is rapid, accurate, and requires virtually no sample preparation.
In conventional NIR spectroscopy, a multivariate statistical model is developed that attempts to correlate subtle changes in the NIR spectrum with known compositional changes determined by standard analytical technology. Once a robust model has been developed, NIR spectroscopic measurements can be substituted for the more time consuming, labor-intensive conventional analytical measurements. To be completely useful, however, a model developed on one spectrometer in the laboratory should be capable of being used on different spectrometers without having to go through the model development all over again with the new instrument. To transfer a model from one spectrometer to another successfully, both instruments must ideally be identical.
In reality, different spectrometers are subtly different. One factor that can have a significant impact on calibration transfer is wavelength accuracy, particularly with dispersive spectrometers. If the wavelength scales of different spectrometers are miscalibrated (as they inevitably are), problems with calibration transfer may occur. One way to avoid this problem is to use a wavelength standard to validate the wavelength scale of the spectrometer. Recently, Busch and co-workers have proposed the use of trichloromethane as a substance with sharp, isolated absorption bands that are suitable for wavelength calibration of spectrometers in the NIR region3. This paper describes an instrumental approach that automatically calibrates the wavelength scale of a dispersive NIR spectrometer without the need for external wavelength calibration materials.
EXPERIMENTALTo permit laser wavelength calibration, a 0.5-mW He—Ne laser (Model 79251, Oriel Corp., Stratford, Conn.), oriented orthogonally to the QTH lamp beam, was positioned between the QTH lamp and the chopper as shown in
A specially constructed sample compartment provided means for double-beam operation of the spectrometer. Radiation emerging from the exit slit of the monochromator was split into two beams by a “polka-dot” beam splitter (Model 38106, Oriel). This beam splitter consisted of a UV-grade fused silica substrate on which was deposited a pattern of reflective aluminum dots, 2.5 mm in diameter, separated by a 3.2 mm center-to-center distance. The beam splitter was positioned so that the laser beam emerging from the exit slit of the monochromator hit the beam splitter exactly on one of the reflective aluminum dots. In this way, radiation from the laser was almost entirely reflected towards the reference PbS detector. The other beam was focused on the sample PbS detector. Signals from the respective PbS detectors were demodulated by two lock-in amplifiers (Model 3962A, Ithaco, Ithaca, N.Y.) before being sent to the data acquisition (DAQ) board of the computer. Overall spectrometer control was accomplished with a program written in LabVIEW™ version 5.1 (National Instruments, Austin, Tex.).
Results and DiscussionPrinciple of Operation. The basic concept behind the laser wavelength calibration system described here is to use a 0.5-mW He—Ne laser to provide known wavelength markers that are recorded simultaneously on the spectrum along with the spectrum of the analyte. These sharp spikes in the spectrum occur at accurately known wavelengths in the spectrum and serve as internal reference points in the spectrum against which the wavelengths of other spectral features may be determined.
The success of the laser wavelength calibration system derives from a combination of factors. First, radiation from a laser is used to provide a small-diameter, highly collimated beam of radiation at an accurately known wavelength (632.8 nm). Radiation from a He—Ne laser is reflected orthogonally by a small mirror so that the laser radiation is co-linear with the beam from the primary QTH lamp. The folding mirror is small (˜3 mm diameter) so that it blocks only a tiny fraction of the primary source beam from the QTH lamp.
Both beams are modulated simultaneously by the rotary chopper and enter the monochromator equipped with the diffraction grating. According to the normal diffraction grating equation5,
mλ=d(sin i±sin θ) (1)
where m is the diffraction order, λ is the wavelength, d is the grating constant, i is the angle of incidence, and θ is the angle of diffraction. According to Eqn. 1, for a given diffraction grating with fixed i and θ, m1λ1=m2λ2.
It is clear from Eqn. 1 that the apparent locations of the 632.8 nm laser line in higher orders should be integral multiples of 632.8 and should, therefore, be spread uniformly throughout the spectrum at m(632.8) nm, where m is the diffraction order in Eqn. 1. Because the optics of diffraction gratings are well known, the positions of the various spikes can be predicted with great accuracy.
Second, the characteristics of the beam splitter in the sample compartment are important. In this study, a so-called “polka-dot” beam splitter was used that had a pattern of reflective aluminum dots deposited on a fused silica substrate. The reflective aluminum dots were 2.5 mm in diameter and were spaced on 3.2 mm centers. For beams larger than 9.5 mm in diameter, the polka-dot pattern of reflective aluminum dots provides a 50/50 split regardless of the angle of incidence. So, for the larger diameter beam of primary source radiation from the QTH lamp, the radiation will be divided approximately equally between the sample and reference beams as desired for double-beam operation. In contrast, by careful placement of the beam splitter, radiation from the collimated laser beam emerging from the monochromator can be made to strike on one of the reflective aluminum dots. In this way, the laser radiation can be almost entirely directed toward the reference PbS detector.
For first-order wavelengths in the NIR region that coincide with the higher diffraction order positions of the 632.8 nm laser line, the intensity of radiation striking the reference detector will go up (i.e., it will consist of radiation from both the QTH lamp and the laser). Since for this spectrometer, absorbance is defined as log(Ireference/Isample), an increase in Ireference will produce an apparent increase in the absorbance at the wavelengths that correspond to higher diffraction orders of 632.8 nm radiation. This will result in absorbance spikes at positions given by m(632.8 nm) in the spectrum, where m is an integer.
Software Control. Overall spectrometer control was accomplished with a program written in LabVIEW version 5.1. For this application, a menu is used to call any one of several virtual instruments (VIs) that allow the user to operate the spectrometer, as well as plot and manipulate data. The complete-plotter VI calls data files, plots spectra, and saves the spectral data with any modifications to another file. Other options include taking derivatives, data smoothing, listing peak locations above an adjustable threshold, and utilization of the laser-calibration option. When the laser-calibration option is selected, the VI corrects the spectrum using parameters calculated in a sub-VI. This sub-VI uses a peak-finding routine available in LabVIEW to locate the laser peaks and determine the parameters used in the complete-plotter VI to calibrate the spectrum.
Equation 2 gives the algorithm that was used for wavelength correction,
λcorrected=(λmeasured+Δ)+β(λmeasured−λ2nd order) (2)
where λcorrected and λmeasured are the corrected and measured values of the wavelength in nanometers, respectively, Δ is 1265.6−λ2nd order, β is 1−α, α is (λ3rd order−λ2nd order)/632.8, and λ2nd order and λ3rd order are the measured values of the second and third order absorbance peaks produced by the laser. The first term in Eqn. 2 shifts the spectrum left or right on the wavelength scale while the second term corrects the dispersion.
Instrument Performance.
It should also be noted that this laser calibration system can be used in two modes of operation. In one mode, the laser peaks are present in the sample spectrum as shown in
To validate the performance of the laser spectrometer, the spectrum of trichloromethane was studied with the laser-corrected spectrometer and a Fourier-transform NIR (FTNIR) spectrometer (Cygnus-25, Mattson Instruments, Madison, Wis.). The wavenumber scale of the FTNIR was calibrated as described previously3.
Table IV shows the wavelength reproducibility obtained from 5 spectra with four trichloromethane bands when laser wavelength calibration is employed. Use of laser wavelength calibration improves the wavelength reproducibility of the spectrometer and reduces the uncertainty in the measured wavelength values to less than 1 nm (average value 0.73 nm).
Table V compares the wavelengths for four bands in the trichloromethane spectrum obtained with the FTNIR spectrometer and the dispersive spectrometer with laser wavelength calibration. The agreement between the two spectrometers is quite good with an average absolute deviation for the four bands of +0.12 nm.
In previous work on the use of trichloromethane as an NIR wavelength standard, a calibrated FTNIR spectrometer was used to determine the wavelength of four bands in the trichloromethane NIR spectrum3. In this study, the wavelengths of these same bands were re-measured with a calibrated FTNIR spectrometer and compared with those obtained with the laser-corrected spectrometer. Table VI summarizes the wavelength values obtained to date for the four trichloromethane bands that have been proposed as wavelength standards for NIR spectroscopy. The results obtained in this study with the FTNIR spectrometer and the laser-corrected dispersive spectrometer are in substantial agreement. Table VI also reports the average values for the wavelengths of the four bands obtained with the two spectrometers used in this study.
CONCLUSIONSIncorporation of a He—Ne laser into a dispersive NIR spectrometer that employs a diffraction grating dispersion system permits wavelength calibration of the instrument based on the known locations of the higher diffraction order positions of the 632.8 nm laser line. Over the spectral range from 1100 to 2000 nm, both the second and third order positions of the 632.8 nm laser line are observed and can be used as markers for wavelength calibration. Agreement between the band positions for chloroform obtained with an FTNIR spectrometer and the dispersive spectrometer with laser wavelength calibration is quite good. The factors that contribute to the proper functioning of the wavelength correction system are: 1) the He—Ne laser emits a sharp, isolated line of known wavelength (632.8 nm); 2) the PbS detector responds to the radiation emitted by the He—Ne laser; 3) the diffraction grating produces multiple orders (out to six) so that the 632.8 nm line appears at known multiples of 632.8 nm in the NIR region; 4) the laser produces a small diameter, collimated beam so that a small mirror can be used to fold the laser radiation into the source beam without obstructing much light from the source beam; 5) a polka dot beam splitter can be arranged so that the laser beam emerging from the monochromator strikes a reflective dot on the beam splitter and is thereby preferentially reflected towards the reference detector; 6) a program written in LabVIEW can be used to perform the wavelength calibration with a simple algorithm.
REFERENCESThe following citations are incorporated by reference herein for details supplementing this application:
- 1. B. G. Osborne, T. Fearn, and P. H. Hindle, Practical NIR Spectroscopy with Applications in Food and Beverage Analysis (Longman Scientific & Technical, Essex, England, 1993).
- 2. Donald A. Burns and Emil W. Ciurszak, Eds., Handbook of Near-Infrared Analysis (Marcel Dekker, New York, 1992)
- 3. Kenneth W. Busch, Olusola Soyemi, Dennis Rabbe, Karalyn Humphrey, Ben Dundee, and Marianna A. Busch, Appl. Spectrosc. 54, 1321 (1999).
- 4. Olusola Soyemi, Dennis Rabbe, Ben Dundee, Marianna A. Busch, and Kenneth W. Busch, Spectrosc. 16(4), 24 (2001).
- 5. Kenneth W. Busch and Marianna A. Busch, Multielement Detection Systems for Spectrochemical Analysis (John Wiley and Sons, New York, 1990) p. 91.
- 6. William L. Wolfe and George J. Zissis, Eds., The Infrared Handbook (Office of Naval Research, Department of the Navy, Washington, D.C., 1985) p. 11-70.
- 7. Kenneth W. Busch, Olusola Soyemi, Dennis Rabbe, and Marianna A. Busch, Appl. Spectrosc. 54, 1759 (2000).
The following citations are incorporated by reference herein for details supplementing this application:
- 1. Donald A. Burns and Emil W. Ciurczak, Handbook of Near-Infrared Analysis (Dekker, New York, 1992).
- 2. H. Martens and T. Næs, Multivariate Calibration (Wiley, New York, 1989).
- T. B. Blank, S. T. Sum, S. D. Brown, and S. L. Monfre, Anal. Chem. 68, 2987 (1996).
- 4. J. J. Workman, P. R. Mobley, B. R. Kowalski, and Ramus Bro, Appl. Spectrosc. Rev. 31, 73.
- 5. J. J. Workman and J. Coates, Spectroscopy 8, 36 (1993).
- 6. Y. Wang, D. J. Veltkamp, and B. R. Kowalski, Anal. Chem. 63, 2750 (1991).
- 7. SRM 1920a. Near Infrared Reflectance Wavelength Standard (National Institute of Standards and Technology, Gaitherburg, Md., 1999)
- 8. SRM 2517: Wavelength Reference Absorption Cell (National Institute of Standards and Technology, Gaitherburg, Md., 1999).
- 9. SRM 2035: Near Infrared Transmission Wavelength Standard from 10,300 cm−1 to 5130 cm−1 (National Institute of Standards and Technology, Gaitherburg, Md., 1999).
- 10. Stephen R. Lowry, Jim Hyatt, and William J. McCarthy, Appl. Spectrosc. 54, 450 (2000),
- 11. Kenneth W. Busch, Olusola Soyemi, Dennis Rabbe, Karalyn Humphrey, Ben Dundee, and Marianna A. Busch, Appl. Spectrosc., 54, 1321 (2000).
- 12. Kenneth W. Busch, Olusola Soyemi, Benn Dundee, Dennis Rabbe, and Marianna A. Busch, Spectrosc., 16(4), 24-33, April, 2001.
- 13. Marianna Busch, Dennis Rabbe, Karalyn Humphrey, and Kenneth W. Busch, “Design and Evaluation of a Near-Infrared Dispersive Spectrometer that uses a He—Ne Laser for Automatic Internal Wavelength Calibration,” 27th Annual Conference of the Federation of Analytical Chemistry and Spectroscopy Societies, Nashville, Tenn., Sep. 25, 2000, Paper No. 108.
- 14. O. Soyemi, Design of Data Acquisition and Analysis Systems in Near-Infrared Spectroscopy: A Virtual Instrument Approach, Ph.D. Dissertation, Baylor University, January, 2000.
- 15. Kenneth W. Busch, Olusola Soyemi, Dennis Rabbe, and Marianna A. Busch, Appl. Spectrosc., 54, 1759 (2000.
- 16. O. Soyemi, Design of Data Acquisition and Analysis Systems in Near-Infrared Spectroscopy: A Virtual Instrument Approach, Ph.D. Dissertation, Baylor-University, January, 2000.
- 17. Christopher G. Wiggenstein, Kirk H. Schulz, and Joe Scott, Rev. Sci. Instrum. 69, 3707 (1998).
- 18. Michel Koenig, Jean M. Boudenne, P. Legriel, A. Legriel, T. Grandpierre, Dimitri Batani, Simone Bossi, Sonia Nicolella, and René Benattar, Rev. Sci. Instrum. 68, 2387 (1997).
- 19. I. W. Kirkman and P. A. Buksh, Rev. Sci. Instrum. 63, 869 (1992).
- 20. Wieslaw J. Stryewski, Rev. Sci. Instrum. 62, 1921 (1991).
- 21. F. J. S. de Viteri and D. Diamond, Analytical Proceedings including Analytical Communications, 31, 229 (1994).
- 22. L. K. Wells, LabVIEW: Student Edition Users' Guide (Prentice Hall, Upper Saddle River, N.J., 1994).
- 23. Peter R. Griffiths and James A. de Haseth, Fourier Transform Infrared Spectrometry, (Wiley, New York, 1986).
Claims
1. A dispersive, diffraction grating NIR spectrometer with the improvement being automatic calibration of the wavelength scale of the instrument without the need for external wavelength calibration means.
2. The spectrometer of claim 1, the improvement further defined as comprising a low power laser at right angles to a source beam of the spectrometer.
3. The spectrometer of claim 1, the improvement further defined as comprising a low power He—Ne laser at right angles to a source beam of the spectrometer.
4. The spectrometer of claim 1, the improvement further defined as comprising a low power He—Ne laser at right angles to a source beam of the spectrometer where the He—Ne laser emits a collimated laser beam with a wavelength of 632.8 nm.
5. The spectrometer of claim 1, the improvement further defined as comprising a mirror to redirect a laser beam so that it is parallel to a source beam.
6. The spectrometer of claim 1, the improvement further defined as comprising a folding mirror to redirect a He—Ne laser beam so that it is parallel to a source beam.
7. The spectrometer of claim 1, the improvement further defined as comprising a beam splitter to divert the majority of a laser beam toward a reference detector.
8. The spectrometer of claim 1, the improvement further defined as comprising a “Polka-dot” beam splitter to divert the majority of a He—Ne laser beam toward a reference detector.
9. The spectrometer of claim 1, the improvement further defined as comprising a beam splitter to divert the majority of a laser beam toward a PbS reference detector.
10. The spectrometer of claim 1, the improvement further defined as comprising a “Polka-dot” beam splitter to divert the majority of a He—Ne laser beam toward a PbS reference detector.
11. The spectrometer of claim 1, the improvement further defined as comprising a PbS sample detector.
12. The spectrometer of claim 1, the improvement further defined as comprising a software routine written in LabVIEW that automatically corrects the wavelength scale of the instrument.
13. The spectrometer of claim 1, the improvement further defined as comprising a software routine written in LabVIEW that automatically corrects the wavelength scale of the instrument from the positions of the 632.8 nm laser line in a spectrum.
14. The spectrometer of claim 1, the improvement further defined as comprising a low power He—Ne laser at a right angles to a source beam where a He—Ne laser emits a collimated laser beam with a wavelength of 632.8 nm; a folding mirror to redirect the He—Ne laser beam so that it is parallel to the source beam; a “polka dot” beam splitter to divert the majority of the laser beam toward a PbS reference detector; a PbS sample detector; a software routine written in LabView that automatically corrects the wavelength scale of the instrument from the positions of the 632.8 nm laser line in a spectrum.
15. A method for improving the calibration capabilities of a NIR spectrometer comprising:
- obtaining an enhanced calibration set of NIR spectra by improving a dispersive, diffraction grating NIR spectrometer so that it automatically calibrates the wavelength scale of the spectrometer without the need for external wavelength calibration means.
16. The method of claim 15 further comprising obtaining a low power laser and placing said laser at right angles to a source beam of the spectrometer.
17. The method of claim 15 further comprising obtaining a low power He—Ne laser and placing said laser at right angles to a source beam of the spectrometer.
18. The method of claim 15 further comprising obtaining a low power He—Ne laser and placing said laser at right angles to a source beam of the spectrometer where the He—Ne laser emits a beam with a wavelength of 632.8 nm.
19. The method of claim 15 further comprising obtaining a mirror and redirecting a laser beam with said mirror so that said mirror beam does not block a source beam from a quartz halogen lamp of the spectrometer and said laser beam is parallel to said source beam.
20. The method of claim 15 further comprising obtaining a folding mirror and redirecting a laser beam with said mirror so that said folding mirror does not block a source beam from a quartz halogen lamp of the spectrometer and said laser beam is parallel to said source beam.
21. The method of claim 15 further comprising obtaining a folding mirror and redirecting a low power collimated He—Ne laser beam with said mirror so that said folding mirror does not block a source beam from a quartz halogen lamp of the spectrometer and said laser beam is parallel to said source beam.
22. The method of claim 15 further comprising obtaining a beam splitter and diverting the majority of a laser beam with said beam splitter toward a reference detector.
23. The method of claim 15 further comprising obtaining a “Polka-dot” beam splitter and diverting the majority of a laser beam with said beam splitter toward a PbS reference detector.
24. The method of claim 15 further comprising obtaining a PbS sample detector and positioning said PbS sample detector such that a source beam light strikes it after passing through a sample cell holder.
25. The method of claim 15 further comprising obtaining and installing a computer software routine that automatically corrects the wavelength scale of the spectrometer.
26. The method of claim 15 further comprising obtaining and installing a software routine written in LabVIEW that automatically corrects the wavelength scale of the spectrometer.
27. The method of claim 15 further comprising obtaining and installing a software routine written in LabVIEW that automatically corrects the wavelength scale of the spectrometer from the positions of 632.8 nm laser line in a spectrum.
28. The method of claim 15 further comprising:
- obtaining a He—Ne low power laser;
- placing said laser at right angles to a source beam of the spectrometer where the He—Ne laser emits a collimated beam with a wavelength of 632.8 nm;
- obtaining a folding mirror; redirecting the low power collimated He—Ne laser beam with said mirror so that said folding mirror does not block the source beam from a quartz halogen lamp of the spectrometer and said laser beam at 632.8 nm is parallel to said source beam;
- obtaining a “Polka-dot” beam splitter; diverting the majority of the laser beam with said “Polka-dot” beam splitter toward a PbS reference detector;
- obtaining a PbS sample detector; positioning said PbS sample detector such that the source beam light strikes it after passing through a sample cell holder;
- obtaining and installing a software routine written in LabVIEW that automatically corrects the wavelength scale of the spectrometer from the position of 632.8 nm laser line in a spectrum.
Type: Application
Filed: Apr 28, 2008
Publication Date: Aug 28, 2008
Inventors: Kenneth W. Busch (Waco, TX), Dennis H. Rabbe (Waco, TX)
Application Number: 12/150,419
International Classification: G01J 3/00 (20060101);