System and Process for Production of Isotopes and Isotope Compositions
Methods for purifying 89Zr are provided, 89Zr compositions are provided, isotope compositions are provided that can include: a radio isotope and a nanoparticle, and methods for radio labeling monoclonal antibodies are provided.
Latest Battelle Memorial Institute Patents:
- Systems and methods for providing virtual system on chip (vSoC) instances
- Calibration of electrode-to-muscle mapping for functional electrical stimulation
- ELECTROLYTES FOR LITHIUM-ION BATTERIES OPERATING AT EXTREME CONDITIONS
- SYSTEMS AND METHODS FOR UTILIZING FOAM FRACTIONATION TO SEPARATE AND ELIMINATE PFAS
- Multimodal inspection system
This application is a divisional of U.S. patent application Ser. No. 15/788,724 filed Oct. 19, 2017, entitled “System and Process for Production of Isotopes and Isotope Compositions”, which claims priority to U.S. Patent Application Ser. No. 62/410,303 filed Oct. 19, 2016, entitled “System and Process for Production of Labeling-Grade Zirconium-89”, the entirety of each of which is incorporated by reference herein.
STATEMENT AS TO RIGHTS TO INVENTIONS MADE UNDER FEDERALLY-SPONSORED RESEARCH AND DEVELOPMENTThis invention was made with Government support under Contract DE-AC0576RL01830 awarded by the U.S. Department of Energy. The Government has certain rights in the invention.
TECHNICAL FIELDThe present disclosure relates to materials, methods, and compositions for use in isotope purification and/or PET isotope production. Particular embodiments provide materials, methods, and compositions for use in 89Zr purification and PET production.
BACKGROUNDAvailability of longer-lived positron emitters has made PET-based imaging of tumors possible by radiolabeling monoclonal antibodies (mAbs), mAb fragments and aptamers, a process referred to as immuno-PET. Immuno-PET combines the high sensitivity and spatial resolution of PET imaging with the antigen specificity of mAbs. Use of 89Zr is gaining tremendous interest in Europe and the U.S. as an immunoPET diagnostic isotope due to its ease of production using monoisotopic (natural) yttrium targets and moderate-to-low energy medical cyclotrons. The long half-life of 89Zr (T1/2=78.4 hr) also enables the potential for off-site isotope production and distribution and for opportunities in new and emerging medical modalities.
Given the high cost of specialty proteins used in immuno-PET diagnostic imaging, it is desirable that binding yields onto these proteins be as high as possible. Another drawback of existing 89Zr purification processing is the high dose rates to personnel that can occur from process scale-up. Accordingly, new processes are needed that address the residual metal contamination for medical applications and personnel dose rates. Embodiments described in the present disclosure address these needs.
Additional advantages and novel features of the present invention will be set forth as follows and will be readily apparent from the descriptions and demonstrations set forth herein. Accordingly, the following descriptions of the present invention should be seen as illustrative of the invention and not as limiting in any way.
SUMMARY OF THE DISCLOSUREMethods for purifying 89Zr are provided with the methods including: loading a hydroxamate resin with a loading solution comprising HCl and 89Zr; and eluting the 89Zr from the resin using an oxalic acid solution having molarity of less than 1. Useful eluting solutions are provided that can include 89Zr and less than 1 M oxalic acid.
Methods for purifying 89Zr can also include: preparing a loading solution comprising 89Zr and HCl; exposing the loading solution to an ion exchange resin; and eluting the 89Zr from resin using an HCl solution having a molarity greater than 0.3.
Methods for purifying 89Zr can also include: loading a first resin with a first loading solution comprising 89Zr; eluting the 89Zr from the first resin using a first eluting solution; loading a second resin with a second loading solution comprising 89Zr; and eluting the 89Zr from the second resin using a second eluting solution.
89Zr compositions are provided that may have bindable metal concentration ([MB]) less than 169 nmole·g−1 or effective specific activity (ESA) of greater than 800 Ci·mmole−1.
Isotope compositions are provided that can include: a radio isotope and a nanoparticle.
Other isotope compositions can include: a radio isotope; a nanoparticle; and a macromolecule.
Methods for radio labeling monoclonal antibodies are provided that can include: preparing a binding conjugate by exposing a ligand to a solution of 89Zr and less than 1 M oxalic acid.
Methods for radio labeling monoclonal antibodies are also provided that can include: exposing monoclonal antibodies to a solution of 89Zr having a have bindable metal concentration ([MB]) less than 169 nmole·g−1 or effective specific activity (ESA) of greater than 800 Ci·mmole−1.
Embodiments of the disclosure are described below with reference to the following accompanying drawings.
This disclosure is submitted in furtherance of the constitutional purposes of the U.S. Patent Laws “to promote the progress of science and useful arts” (Article 1, Section 8).
The materials, methods, and compositions of the present disclosure will be described with reference to accompanying
Referring first to
In accordance with example implementations, methods for purifying the isotope 89Zr, can include loading a hydroxamate resin material, in a column, with a loading solution comprising HCl and 89Zr. This solution can be directly obtained with the dissolution of yttrium. Example irradiated yttrium can include, but is not limited to, cyclotron bombarded Y foil, for example. Some metals of the solution, M1 and Mz, may be bound to the resin, and much of the remainder of the loaded solution removed from the resin.
In accordance with example implementations, the resin can be exposed to an eluting solution, in this case a solution of oxalic acid having a molarity less than 1. The oxalic acid solution (H2C2O4) can be between 0.3 and less than 1 M, or 0.8 and less than 1 M. In accordance with example implementations, a solution can be obtained during elution in system 20 that includes 89Zr and less than 1 M oxalic acid. For example, this results in the need for less buffering agents to be added to the 89Zr product fraction prior to mAb labeling. In accordance with example implementations, eluted 89Zr product fractions demonstrated chemical recoveries from irradiated Y foil targets, with 1 mL product volumes yielding 89±2% of the elution peak activity and 84±2% of the 89Zr from the irradiated target.
Referring to
The fluidic system has been used to demonstrate the performance of the 89Zr purification method commonly used in the field. However, the method has been modified to utilize a significantly less concentrated oxalic acid eluent solution (0.8 M), which primarily serves to minimize the quantity of buffering agents required to be added prior to labeling the 89Zr product. The concentration of oxalic acid used in the field (1 M) can be close to saturation at laboratory temperatures; the solubility limit of H2C2O4 in water at 20° C. is 95.2 g·kg−1 (1.06 mol·kg−1), or a 8.69% (w/w) mass fraction. The methods and systems may reduce the amount of metal contaminants introduced into the product. With reference to
Optima™ grade hydrochloric acid (Fisher Scientific, Waltham, Mass.) was used for Y target dissolution and column washes, and hydrogen peroxide (29-32%, Sigma-Aldrich) was used in the Y target dissolution. TraceSELECT® oxalic acid dihydrate (>99.9999%, Sigma-Aldrich, St. Louis, Mo.) was used for 89Zr column elutions. All solutions were prepared from deionized water (18.3 MΩ·cm) using a Barnstead Nanopure Diamond water purification system (Dubuque, Iowa). Hydroxamate ligand was bound onto Accell Plus CM weak cation-exchanger (37-55 μm, Waters Corp., Milford, Mass.) by the method of Verel et al. The hydroxamate column was prepared using 100 mg resin packed into a disposable 1 cc SPE column using 2 mm thick polyethylene (PE) frits (Supelco, Bellefonte, Pa.). With an ID of 5.6 mm, the resin bed had a height of ˜1.0 cm.
Zirconium-88 (88Zr) and yttrium-88 (88Y) radiotracers (tracers), with half-lives of 83.4 and 106.6 days, respectively, were purchased through the Department of Energy Office of Science National Isotope Development Center (NIDC). The 88Zr was periodically purified of in-grown 88Y using an anion exchange separation method, whereby 88Zr was loaded onto a ˜0.8 cc AG MP-1M column (Bio-Rad Laboratories, Inc., Hercules, Calif.) in concentrated (conc.) HCl (88Y is unretained). Then, after a conc. HCl column wash, the 88Zr was eluted in 6 M HCl. 89Zr was obtained from the University of Washington (UW) PET radiochemistry center as an irradiated Y foil of 0.25 mm thickness (Alfa Aesar, Ward Hill, Mass.). Following an initial decay period of ˜2 h, the 89Zr target activity was determined by a dose calibrator and corrected to activity at end of bombardment (EOB). Upon delivery to PNNL, this activity was verified using a high purity germanium (HPGe) detector (Ortec, Oak Ridge, Tenn.) calibrated with NIST-traceable standards, with energy and efficiency checks performed daily.
Referring to
As shown in
The Y foil was dissolved in a 50 mL 2-neck round-bottom reaction flask (RF) that contained a magnetic stir bar. A 4-channel Sci-Q 400 peristaltic pump (PP) (Watson-Marlow, Wilmington, Mass.) was configured to deliver the Y foil dissolving solution via a tube inserted through the side neck of the RF using pump channel 1 (not shown). Channel 2 of the PP was used to withdraw the dissolved Y foil solution via a polyetherimide (ULTEM™) sipper tube inserted through the central neck. Downstream of the PP, an in-line filter (17 mm diameter, 0.45 μm PES syringe filter (Thermo Scientific) was connected to the delivery line using PEEK luer to ¼-28 couplers (Upchurch). The dissolved target solution delivery line was then routed to the 6-port 2-position V. In position 1, the V directed the PP flow to a waste line, and in position 2, the PP connected to the column. Inversely, the SP access to the valve's waste and column delivery ports were positions 2 and 1, respectively.
The fluidic system reagent delivery protocol for purifying 89Zr on a hydroxamate column is summarized in Table 1.
The column is first conditioned with 2 M HCl by the syringe pump (SP) via the valve (V), which is initially set to position 1. The irradiated Y foil (typically 150-250 mg) is placed in the 2-neck round-bottom flask. The peristaltic pump (PP) is engaged to deliver 6 mL 2 M HCl/0.1 M H2O2 to the side neck of the flask at a rate of ˜6 mL/min. The solution is stirred during dissolution with a magnetic stir bar to minimize frothing and spatter inside the flask. The solution is allowed to cool with stirring for 10-20 min, followed by 5-10 min of settling prior to initiating column load. Next, the dissolved foil solution is delivered to the column at 0.5-0.6 mL/min by the PP through an in-line filter (F) to remove residual solids, and then through the V (toggled to position 2). Once the solution is transferred onto the column, the V is returned to position 1, where the SP is once again connected to the column. The SP sequentially delivers 10 mL of 2 M HCl and 10 mL of H2O washes to the column at a 1 mL/min flow rate. At the completion of the H2O wash, and prior to the 89Zr elution step, the SP pushes air through the column delivery line until all fluid is purged from the column and detector/fraction collector delivery lines. Zirconium-89 can be eluted using 5 mL of H2C2O4, for example. The fraction collector is used to separate the elution volumes so only the first ˜1-1.5 mL of the elution volume can be used for subsequent evaluation of product labeling performance.
The 89Zr purification process is 1) fluidically automated, 2) provides near-real time monitoring of column effluents, and 3) uses a significantly lower concentration of H2C2O4 (0.8 M) eluent than that reported previously. A schematic of the fluidic device is shown in
On-line detector traces for a hydroxamate resin column separation of 89Zr from an irradiated Y foil target are shown in
Five Y foils were irradiated and then processed as per the described hydroxamate resin column method. Table 2 provides a summary of the 89Zr activities (at EOB) and masses for each foil, as well as the chemical separation performance of 89Zr from the dissolved foils.
Irradiated foils 1-3 were originally ˜150-180 mg, but the mass was augmented with the addition of unirradiated Y foil to ˜250 mg in order to assure the robustness of the method. Foils 4 and 5 received no additional Y foil. For each 89Zr purification, all column effluents were collected in order to determine the fraction of 89Zr activity at each step. The total elute fraction contained 99.0±0.9% of the column-loaded activity. However, some 89Zr activity was lost to the system, either in the form of residues remaining from the foil dissolution (activity remaining in the reaction flask and in-line membrane filter) or as activity trapped in the column. Residual 89Zr activity was calculated to be 4±1% of the original foil activity. Considering the 89Zr activity lost to the system, the total 89Zr yield in the elution fractions was 95±2% of the 89Zr present in the irradiated Y foil.
The elution profiles for four replicate 89Zr purification runs are shown in
With reference to
In accordance with example implementations, three strongly basic anion exchange resins were evaluated. Each had quaternary amine functional groups with different polymer supports. The supports ranged from macroporous and 10% cross-linked styrene-divinylbenzene copolymer to a hydroxylated methacrylic polymer. The resins were evaluated for their ability to load and elute Zr in neat solutions and solutions containing high concentrations of dissolved Y. Additionally, the method had to ensure removal of Fe (a metal that competes for antibody labeling sites for example) from the Zr product. In accordance with example implementations, eluent solutions are provided to remove 89Zr selectively to Fe(III). Accordingly, after a dissolved irradiated Y foil was loaded and subsequently washed, both 89Zr and Fe may remain on the resin. Eluting solutions are provided that elute 89Zr while leaving Fe(III) on the resin.
The 89Zr purification method involves cyclotron bombarded Y foil dissolution, loading of 89Zr from the dissolved target onto an anion exchange column, column wash, and 89Zr elution. This sequence of steps was accomplished with a prototype automated fluidic platform (
TraceMetal and Optima grade hydrochloric acid (34-37% HCl, Fisher Scientific) working stock solutions were prepared as-is or as dilutions into deionized water (18.3 MΩ·cm) using a Barnstead E-Pure water purification system (Dubuque, Iowa). Method development work and cyclotron bombarded Y foil processing used TraceMetal and Optima grade acids, respectively. Sodium fluoride (99.99+%) and hydrogen peroxide (30%) were purchased from Sigma-Aldrich (Milwaukee, Wis.). Several ion exchange resins were evaluated: AG MP-1M (Cr form, 200-400 mesh) and AG 1-X10 (Cl− form, 100-200 mesh) strongly basic anion exchangers, each on styrene-divinylbenzene co-polymer (Bio-Rad Life Science, Hercules, Calif.), and Toyopearl QAE-550C strongly basic anion exchanger (Cl− form, 50-150 μm) on hydroxylated methacrylic polymer (TOSOH Bioscience, King of Prussia, Pa.).
As depicted in
Two automated fluid delivery protocols were prepared for optimization of 88Zr tracer separation methods on ion exchange columns. For initial column performance studies, 88Zr tracer was introduced to the column using a 0.29 mL sample loop configured within a 6-port 2-position valve (V1). A description of the protocol is shown in Table 3, and the system is illustrated in
The SP was programmed to deliver reagents at 0.75 mL/min. With V1 in position 1, the sample loop (SL) was isolated, and the SP delivered solutions directly to the column to be conditioned with conc. HCl. Next, the 88Zr was loaded onto the column by toggling V1 to position 2, whereby a 5 mL aliquot of load solution from the SP pushed the 88Zr-bearing solution to the column and subsequently washed the column. Next, 7.5 mL of 88Zr eluent solution (varying between 0.1 and 9 M HCl) was delivered to the column, followed by 4 mL of strip solution to ensure all 88Zr was removed from the column and the tubing.
Once the optimal separation media and 88Zr tracer eluent solution was selected, a second fluid delivery protocol was prepared, and is summarized in Table 4.
The hardware was reconfigured to that shown in
Batch contact experiments were performed in order to determine the distribution coefficients (Kd) of Zr and Y (as 88Zr and 88Y) on two anion exchange resins across a range of HCl concentrations (0.1 to ˜12 M). AG MP-1M (MP-1M) and Toyopearl QAE-550C (QAE-550C) were evaluated. A Kd map for Zr on each resin as a function of HCl concentration is shown in
The Kd map shows low Kd at HCl concentrations below ˜9 M, and steeply rising Kd as the HCl concentration approaches ˜12 M. As HCl concentration increases, Zr chloride complexes transition from ZrCl4→ZrCl5−1→ZrCl6−2, the latter of which is most strongly retained on the quaternary amine group of the resins.
Both MP-1M and QAE-550C were evaluated for their performance in loading and eluting 88Zr tracer from a 4.6×50 mm column. The fluidic platform was configured as shown in
The activity in the eluent fraction from the MP-1M resin is poor at low concentrations and increases to a maximum at 6 M HCl (0.86±0.04 elute/0.14±0.05 strip), above which it decreases slightly, and ultimately falls to near zero at 9 M HCl.
The incomplete elution of 88Zr tracer on the macroporous MP-1M prompted the evaluation of the more traditional gel-type anion exchange resin AG 1-X10 (Bio-Rad) under the same conditions. However, the elution profile at 6 M HCl was virtually identical to MP-1M at the same acid strength (0.81 elute/0.13 strip).
The experiments were repeated with the column packed with QAE-550C across a range of HCl concentrations QAE-550C exhibited significantly improved chromatographic performance, with a substantially higher elute fraction across all HCl concentrations tested. 88Zr tracer elution recoveries were observed between HCl molarities of 0.3 (˜92%) to 9 (˜97%). Eluent condition ranges can also be between 2 and 6 M HCl (˜97%) (Table 5).
Once the optimal eluent solutions were obtained for MP-1M (6 M HCl) and QAE-550C (2 M to 6 M HCl), the automated protocol was repeated with 15.9 kBq 88Y tracer in conc. HCl loaded into the 0.29 mL sample loop. Yttrium-88 tracer was observed to be thoroughly removed from the column during the 88Zr during the load/wash step. Using a 6 M HCl elute solution, the MP-1M and QAE-550C 88Y distributions for the load/wash—elute—strip fractions were 0.999/0.001/0.000 and 0.998/0.002/0.000, respectively.
Despite MP-1M resin's ability to better retain Zr in the presence of high concentrations of Y, its chromatographic performance was poor, with incomplete elution (vide supra). An approach was therefore sought to improve the column's Zr elution performance. Zr produces chlorofluoro complexes in the general form (ZrCliFj)(4-i-j). The presence of low concentrations of HF across a range of HCl concentrations can have a profound effect on the anion exchange behavior of Zr when compared to the behavior of Zr in HCl alone. Specifically, the presence of HF prevents the steep rise in Zr affinity for anion exchange resin at high HCl concentrations. This is likely caused by the formation of chlorofluoro complexes with high negative charges.
An experimental matrix was performed to determine how effective the fluoride ion (F−) was in enhancing the Zr elution performance. Solution combinations of HCl and fluoride ion (as NaF) were prepared with concentrations ranging between 1 and 6 M (1, 2, 3, 4, 6 M) and 0 and 1 mM (0, 0.1, 0.33, 1.0 mM), respectively. Each solution in this matrix was evaluated for its ability to efficiently elute 88Zr tracer from a pre-loaded column.
Simulated dissolved irradiated Y targets were prepared using non-irradiated Y metal spiked with 88Zr or 88Y tracers following metal dissolution and solution filtration. One-half gram of Y (Alfa Aesar pieces) was utilized in order to robustly verify method performance, as an actual Y target mass would be expected to be significantly lower than this. Two allotments of 0.5 g Y metal were dissolved in 20 mL conc. HCl each using a 50 mL round-bottom flask (with magnetic stirring). After allowing the solution to cool, the solution was filtered (0.45 μm PES, Thermo Scientific) and spiked to ˜1 kBq/mL 88Zr (solution 1) and 6.5 kBq/mL 88Y (solution 2) tracers, respectively. The fluidic platform was configured as shown in
The peristaltic pump was operated until the entire sample volume, plus rinses (0.5 mL×2), had been delivered to the column (the rapid drop in 88Y tracer signal at the end of the load is air pushed through the detection coil). The small 88Y peak shown in the wash segment (arrow) represents residual 88Y droplets removed from the fluid delivery lines plus removal of residual sample load solution trapped in the interstitial spaces of the column. The 88Y tracer, used as an indicator of the dissolved Y metal in the simulated dissolved target solution, shows virtually complete removal of Y after just a couple of milliliters of wash solution. Of 130 kBq 88Y tracer in 0.5 g dissolved Y metal delivered to the column, <6 Bq were observed in the subsequent Zr elute fraction (signal was below the gamma detector detection limit). This activity is equivalent to <23 μg Y, and represents a Y decontamination factor (DF) of >22,000.
An additional set of experiments was performed in order to evaluate the Zr loading performance of the MP-1M column across a range of Y concentrations and load volumes in conc. HCl. Table 6 summarizes the load conditions and the observed Zr breakthrough levels.
Despite the Zr Kd dependence on Y concentration (
Six separation protocol variations, each representing only a change in the eluent solution delivery, were run using the Table 4 protocol. The load solution contained 0.5 g dissolved Y metal and 88Zr tracer in 20 mL conc. HCl. The elute (reduced to 5 mL) and strip portion of the detector traces for each run are shown in
Set A shows FF and RF elute and strip traces for 88Zr with an eluent solution of 6 M HCl. By the end of the FF elution volume, only 91.4% of the retained 88Zr was removed from the column. For the 6 M HCl RF trace, 98.5% of the 88Zr was recovered, with 97% being removed in 2.9 mL. Set B shows FF and RF traces when the eluent solution was 6 M HCl+0.33 mM NaF. The presence of fluoride ion causes a discernible sharpening of the elution peak for the FF condition, with the RF condition showing only a slightly enhanced elution profile. The FF condition resulted in 98.9% of the retained 88Zr tracer being removed in 5 mL, with 97% removal within 2.9 mL, while the RF condition yielded 99.4% recovery in 5 mL and 97% removal within 2.4 mL of the eluent solution. Set C shows the same for an eluent delivery of 6 M HCl+1.0 mM NaF. Here, the increased complexing strength of the higher fluoride ion concentration results in almost indistinguishable 88Zr tracer elution profiles between FF and RF. The 5 mL elution recoveries are 99.5% and 99.3%, respectively, while 97% is recovered in 2.4 mL for both flow direction conditions. Based on these observed elution profiles, it was concluded that a solution of 6 M HCl+0.33 mM NaF provided essentially equivalent elution profiles of Zr from the MP-1M column, regardless of eluent solution flow direction.
The test Y foils (Alfa Aesar, 0.25 mm thickness) were cyclotron irradiated to produce 89Zr activities ranging between ˜0.2 and 1.6 mCi. The foils were dissolved in conc. HCl, and the resulting solutions delivered to the column as described above; the solutions had Y concentrations between ˜9 and ˜19 mg/mL.
The fluid delivery protocol outlined in Table 4 was implemented to perform the 89Zr/Y separation. The 89Zr was eluted with 6 M HCl+0.33 mM NaF prior to delivery of the column strip solution. Table 7 provides a summary of the performance from eight 89Zr purification runs performed under three conditions: 1) RF elution direction with no H2O2 added to the dissolving solution; 2) RF direction with H2O2 added to the dissolving solution; and 3) FF elution direction with H2O2 added to the dissolving solution. The table lists the column effluent distributions for 89Zr determined by off-line measurement of all collected fractions.
In order to account for the distribution of all 89Zr in the irradiated foil (not only 89Zr delivered to the column), post-run fluidic system washes were performed, and the solutions were counted off-line. These washes included the round bottom flask and the dissolved target delivery line and filter. Additionally, the washed in-line filter and column were sent to the counting lab to determine 89Zr activity in the trapped solids. The last two columns of Table 7 present 1) the fraction of 89Zr lost in the fluidic system as droplets and as insoluble particulate matter (the primary source of 89Zr activity loss), and 2) the 89Zr column elute fraction compared to the total 89Zr present in the irradiated foil.
The conc. HCl used to dissolve the first irradiated targets did not contain H2O2 (foils 1 and 2, RF direction). These were small targets, each ˜200 mg. Surprisingly, a significant fraction (8-10%) of 89Zr was lost to the column during the sample load, and ˜6% was recovered in the strip fraction. Approximately 10% of the 89Zr activity was found as insoluble residue. Combined, this resulted in a ˜76% 89Zr eluent recovery from the overall foil activity. The poor 89Zr yield in the elute solution had not been observed during simulated dissolved Y target studies with 88Zr spike, and is perhaps evidence of a persistent non-anionic Zr complex or colloidal solid formed during cyclotron irradiation or during target dissolution (e.g., zirconium oxide or zirconyl species).
The effect of H2O2 in the HCl dissolver solution on 89Zr recovery was therefore evaluated. Foils 3-5 (RF direction) were processed as above, but the conc. HCl reagent was spiked to 0.1 M H2O2 immediately prior to target dissolution. Despite the increased masses of these foils (between ˜240 and ˜370 mg Y), the amount of 89Zr lost in the load dropped significantly to <1%. Additionally, the amount of 89Zr in the strip fraction dropped from ˜6% to 0.7±0.4%. The 89Zr elution fraction recovery now reached 98.9±0.6% of column-delivered effluents, and 95±3% of the overall foil activity. Foils 6-8, performed with H2O2 in the dissolver solution and the preferred FF column elute direction, performed equally as well, with 97±1% of the column delivered activity, and 96±2% of the overall foil activity, found in the 89Zr elute fraction.
The 89Zr product fractions from Foils 6-8 were analyzed for Y, Fe, and Zr by inductively coupled plasma-mass spectrometry. It was determined that the three product fractions contained 0.74±0.38 μg Y from Y target masses that averaged 0.23±0.02 g. This represents a Y decontamination factor of ˜3×105. Iron in the 89Zr product fractions was below the detection limit of the ICP-MS. Therefore, the mass of Fe present could only be reported as <0.16 μg. Based on the certificates of analysis (COA) for the Y metal foils used in this study, the irradiated foils would be expected to contain 66±38 μg Fe; therefore, a conservative Fe decontamination factor of >410 is reported. However, given the excellent separation of Zr from Fe using this method (
Referring next to
In accordance with additional implementations and with reference to
Accordingly, methods for purifying 89Zr are provided with the methods including loading a first resin with a first loading solution. This first loading solution can include dissolved Y foil that includes 89Zr, for example. The method can then include washing the first resin and then eluting the 89Zr from the first resin using a first eluting solution. In the case where the first resin is an ion-exchange resin, the elution can be performed with an HCl solution or an HCl and fluoride solution. The method can then include loading the second resin with a second loading solution comprising 89Zr. In some embodiments, the loading solution for the second resin can be 89Zr in HCl solution. In some embodiments, the loading solution for the second resin can be 89Zr in HCl obtained from the eluting solution from the first resin. The 89Zr on the second resin can be eluted with a second eluting solution such as oxalic acid. In accordance with example implementations, the first resin can be an ion exchange resin and the second resin is a hydroxamate resin or HDEHP resin.
In accordance with example implementations, TraceMetal and Optima grade hydrochloric acid (34-37% HCl, Fisher Scientific) working stock solutions were prepared as-is or as dilutions into deionized water (18.3 MΩ·cm) using a Barnstead E-Pure water purification system (Dubuque, Iowa). Method development work and irradiated Y foil processing utilized TraceMetal and Optima grade acids, respectively. Sodium fluoride (99.99+%), hydrogen peroxide (30%), and TraceSELECT® oxalic acid dihydrate (>99.9999%) were purchased from Sigma-Aldrich (St. Louis, Mo.). AG MP-1M strongly basic anion exchange resin (CI− form, 200-400 mesh), with quaternary amine functional groups on a macroporous styrene divinylbenzene co-polymer (Bio-Rad Life Sciences, Hercules, Calif.) was used for the primary column. Hydroxamate ligand was synthesized onto Accell Plus CM weak cation-exchanger (37-55 μm, Waters Corp., Milford, Mass.) and used for the secondary column. The MP-1M column had internal dimensions of 4.6×50 mm (0.83 cc internal volume) (OmegaChrom column, Upchurch Scientific, Oak Harbor, Wash.), and used frits from IsoLab, Inc. (Akron, Ohio). The hydroxamate column was constructed from a 2 cm long, ˜27 μL internal volume, internally tapered mini-column with PTFE frits (GlobalFIA, Fox Island, Wash.).
Two fluidic system configurations were utilized for this work. Initially, a simple fluidic system was assembled to evaluate the performance of a second resin, in this case a single hydroxamate microcolumn, as shown in
The second system was configured for an automated tandem column purification of 89Zr from irradiated Y foils. It was comprised of two digital syringe pumps (SP1 & SP2), each with 8-position distribution valves. The SP1 and SP2 syringe displacement volumes were 10 mL and 2.5 mL (Flex Fluidics), respectively. A 6-port 2-position valve (V1) and two 4-port 2-position valves (V2 & V3) were Valco Cheminert valves with microelectric actuators. The pumps and valves were assembled into separate. Once again, back-flow preventers were installed in the SP1 and SP2 reagent delivery lines where they crossed the radiological boundary. In lieu of a SL in V1, the second system incorporated the ability to push a dissolved irradiated target solution from a 50 mL round-bottom flask via a peristaltic pump (PP) to the primary column when V1 was in position 2; SP1 could deliver reagent to the C1 when V1 was in position 1. The target dissolution chamber and dissolved target delivery system have been described herein.
Two automated fluid delivery protocols were prepared in order evaluate the load/wash-elute performance of 88Zr tracer using column-based separations. The first protocol was designed to evaluate the performance of a hydroxamate microcolumn in isolation from the primary anion exchange column. The steps of the automated protocol are presented in Table 8.
Solutions were aspirated into the 2.5 mL syringe and delivered at 0.2 mL/min. With V1 in position 1, the SP could deliver the conditioning solution to the column, while 88Zr or 88Y tracer spiked samples remained isolated in the loop. The sample was loaded onto the column, along with 2.5 mL wash solution, after toggling V1 to position 2. Finally, 2.5 mL of 88Zr tracer eluent solution (H2C2O4) was delivered to the column after V1 had returned to position 1.
Once ideal load/wash-elute conditions were obtained for the hydroxamate microcolumn, a tandem column purification method for 89Zr from dissolved Y foil was developed. The separation protocol is presented in Table 9, and the fluidic system shown in
Initially, the primary (C1) and secondary (C2) columns were conditioned with conc. HCl and 2 M HCl using SP1 and SP2, respectively. Next, the primary column was loaded with the 89Zr from the dissolved Y metal target using the PP. Following sample load, the primary column was washed with additional conc. HCl using SP1 to remove all traces of dissolved Y. The 89Zr on the primary column was then transferred to the secondary column by toggling V2 to position 2 and delivering 5 mL of 6 M HCl+0.33 mM NaF with SP1 to both columns. With all 89Zr activity now on the secondary column, SP2 commenced to deliver 2 M HCl and H2O wash solutions, respectively. Finally, the 89Zr was eluted from the secondary column, using H2C2O4, to 89Zr product collection vessels (CV) with V3 positioned to position 2. While in route from the microcolumn, the 89Zr elution profile was monitored by the in-line gamma detector.
As herein described, a primary column (MP-1M) eluent solution of 6 M HCl+0.33 mM NaF could be utilized as a transfer agent to a secondary microcolumn packed with hydroxamate resin. Additionally, it demonstrated how a solution of H2C2O4 could be utilized to efficiently elute Zr from the hydroxamate microcolumn in a volume (˜0.29 mL) that is significantly less than that specified for the traditional method (typically reported as ˜1 mL or greater). The next step was to develop a fully automated tandem column Zr separation that would combine the optimized method for an MP-1M column with the optimized hydroxamate microcolumn method as primary and secondary purification stages, respectively.
The hydroxamate column can be traditionally conditioned in 2 M HCl, after which the dissolved irradiated Y target (in ˜2 M HCl±H2O2) is loaded to adsorb the 89Zr. The column is subsequently washed with 2 M HCl to remove residual Y ions, and is then washed with H2O to remove residual HCl. The 89Zr is traditionally eluted with 1 M H2C2O4, and the first 1 mL eluent fraction is typically collected for subsequent radiolabeling.
Given the familiarity and reliability of this separation method, it was replicated to the extent possible in the latter steps of the MP-1M→hydroxamate resin tandem column separation scheme. Table 9 summarizes the tandem column separation protocol that was employed using the fluidic system shown in
With reference to Table 10 and
On-line detector traces for a tandem column separation of 89Zr from an irradiated Y foil target are shown in
A total of seven cyclotron bombarded yttrium metal foils were cyclotron irradiated and processed as per the described tandem column method. Table 11 provides a summary of the 89Zr foil activities (at EOB) and masses, as well as the chemical separation performance of 89Zr from the dissolved foils.
Irradiated foils 1-5 were originally between 78 and 180 mg, but the mass was augmented with the addition of non-irradiated Y foil to ˜250 mg in order to assure that the hydroxamate microcolumn performance was robust. Foils 6 and 7 received no additional Y foil. Irradiated Y foils were dissolved with a solution of 20 mL conc. HCl+0.1 M H2O2. For each of the seven 89Zr purifications, all column effluents were collected in order to determine the fraction of 89Zr activity at each step. The hydroxamate microcolumn elute fraction contained 96.8±1.4% of the column-recovered activity. However, some 89Zr activity was lost to the system, either in the form of residues remaining from the foil dissolution (activity remaining in the reaction flask and in-line membrane filter) or as activity trapped in the two columns. Residual 89Zr activity was calculated to be 1.9±0.4% of the original foil activity. Considering the 89Zr activity lost to the system, the total 89Zr yield in the elution fractions was 95.1±1.3% of the total 89Zr foil activity.
The elution profiles for 89Zr from the hydroxamate microcolumn are shown for four replicate runs in
In order to establish a purity measurement specific for 89Zr compositions analysis methods were developed. For example, Deferoxamine is a bacterial siderophore that is the principal conjugate used to radiolabel proteins with 89Zr in immunoPET applications. However, metal contaminants that may persist in the 89Zr product fraction following hydroxamate column purification from the irradiated foil may decrease 89Zr binding yields. These metals include Zr, Fe, Ga, Al, etc. In general, the concentration of these potentially interfering metals in the 89Zr product fractions are dependent on the impurity levels in the Y target metal, the target mass required for irradiation, and the decontamination factors made possible by the chemical purification method employed. Commercially available deferoxamine mesylate (Dfo-m) was used to assess the quality of the 89Zr product fractions that were collected from the automated hydroxamate resin column purification method described above.
The following reagents were used for 89Zr labeling of deferoxamine mesylate (Dfo-m): Deferoxamine mesylate salt was obtained from Santa Cruz Biochemicals (≥98%) and Sigma-Aldrich (99.0%). Dimethyl sulfoxide (DMSO, ACS Reagent ≥99.9%), sodium carbonate (TraceSELECT®), sodium chloride (TraceSELECT®), and 4-(2-Hydroxyethyl)piperazine-1-ethanesulfonic acid (HEPES, 99.5%) were supplied by Sigma-Aldrich. 2,5-Dihydroxybenzoic acid (gentisic acid, 99%) was supplied by Acros Organics (New Jersey).
Parameters that were evaluated in this assessment included bindable metal concentration ([MB]) and effective specific activity (ESA). Two 89Zr/Dfo-m binding solution matrices were used to determine these parameters for 89Zr product fractions from each of three irradiated Y targets purified by the method described above.
Method 1 employed a simple solution preparation that consisted of only H2C2O4, Na2CO3 and H2O in an attempt to minimize the addition of competitive metals from impurities in reagents (Table 12); Method 2 used a more complex solution mixture that incorporated saline, buffers (Na2CO3 and HEPES), and gentisic acid (an anti-oxidant), similar to that reported by Verel et al. Both methods resulted in solutions with a pH of ˜7.
Following solution mixing and incubation, the 89Zr binding fraction was determined with the use of an aminopropyl-based anion exchange cartridge. Zirconium-89 oxalate (likely 89Zr (C2O4)4−4) that was successfully transchelated to Dfo-m (89Zr/Dfo-m) was not retained on the resin, while non-transchelated 89Zr(C2O4)4−4 was retained. The 89Zr/Dfo-m binding fraction was plotted as a function of added Dfo-m concentration, and a sigmoidal curve was fitted (
Table 13 provides details of the Dfo-m transchelation experiments for Methods 1 and 2 using 89Zr product fractions obtained from Foils 3-5 using the hydroxamate resin column-based 89Zr purification method.
Method 1 employed a range of 89Zr product volume fractions ranging from 1.3 to 3.7%; H2C2O4 and Na2CO3 additions varied, but their concentrations were equimolar so that complete solution neutralization was achieved, and solutions were diluted with H2O. Whereas a single 89Zr/Dfo-m transchelation curve was generated for Foil 3, duplicate and triplicate curves of different transchelation solution mixtures were generated for Foils 4 and 5, respectively. The mean [MB] value was determined to be 115±27 nmole·g−1, with an ESA of 1180±190 Ci·mmole−1.
In contrast, Method 2 employed a consistent 10% volume fraction between 89Zr product and the buffered Dfo-m transchelation mixture. The only H2C2O4 added to the experiment was from the 89Zr product, which resulted in a final H2C2O4 concentration of 80 mM. Across the three foils evaluated, Method 2 provided consistent [MB] values averaging 43±7 nmole·g−1. ESA was calculated as 2940±600 Ci·mmole−1. The [MB] and ESA results for Methods 1 and 2 were within a factor of ˜2.5 of each other, thus providing good agreement.
While the ESA term is used ubiquitously in reporting medical isotope quality, it cannot be used to compare isotope products of different activity and/or originating target mass; it is simply the ratio of isotope activity to bindable metals content (MB). The term [MB] expresses the concentration of metals present in the isotope product fraction that compete for binding to the labeling chelate (MB), normalized to the mass of target material from which the isotope is produced. Isotope activity is not considered, because the radioactive isotope produced within the target (89Zr=25 nmol/Ci) constitutes an insignificant portion of the competing metal presence (e.g., elemental Zr, Fe, etc.) within the purified product. Consideration of target mass is important, as the target metal (in this case, Y foil) is likely the greatest source of metal impurities introduced into the separation process. Since any given chemical purification process is capable of a finite decontamination of non-Zr metal impurities from the target metal, the use of a lower-mass target should yield a lower MB. Additionally, a lower-mass target will contain less elemental Zr contamination in the 89Zr product, which should likewise assure a lower MB. Normalizing MB to the mass of target from which the isotope was produced therefore negates target mass differences and allows for a direct comparison of a purification method's efficacy. This assumes, of course, that the metal contaminant concentrations in high purity Y targets are approximately consistent.
Performance of the tandem column separation method was demonstrated by processing seven cyclotron irradiated Y foil targets. The 89Zr elution from the secondary column represented 96.8±1.4% of the total activity collected from the dual column process, and 95.1±1.3% of the total activity present in the irradiated foil. The evaluation of replicate 89Zr elution traces lead to the determination that 90% of the elution activity is found in the first 0.29±0.06 mL of H2C2O4 eluent solution; this elution fraction represents 86±5% of the activity in the irradiated Y foil. This represents an 89Zr product that is times more concentrated than that reported by this team and others when performing the traditional single hydroxamate column method. The combination of an increased 89Zr product concentration (reduced product volume) and reduced H2C2O4 concentration should result in improvements in the labeling performance of 89Zr onto conjugated mAbs.
In accordance with the materials and methods described herein, 89Zr compositions are provided that can have bindable metal concentration ([MB]) and effective specific activity (ESA), respectively of less than 169 nmole·g−1 and/or greater than 800 Ci·mmole−1.
In accordance with example implementations, these compositions can be used as starting materials for methods for radio labeling monoclonal antibodies, other proteins, aptamers, etc.
Methods are also provided for binding 89Zr to a macromolecule. The methods can include binding the 89Zr to a nanoparticle and then subsequently binding the 89Zr-nanoparticle to a macromolecule such as protein including immunoglobulin(s). Accordingly, isotope compositions are provided that can include a radio isotope and a nanoparticle and/or a macromolecule. In accordance with example implementations, the isotope composition of claim 25 wherein the radio isotope is one or more of 89Zr, 86Y, and/or 90Y.
In accordance with example implementations, batch contact studies were performed to evaluate the uptake of both 88Zr and 88Y (both in the chloride form) onto magnetic iron oxide nanoparticles. Chemically unmodified, commercially available Fe3O4 (“FeO”, ˜27nm) and in-house synthesized manganese-doped Fe3O4 (“MnFeO”), were used. The distribution coefficient values are shown in
The nanoparticle uptake of highly purified 89Zr was obtained directly from the 0.8 M oxalic acid elution matrix. Each of these methods result in 89Zr product fractions in the oxalate form (likely as Zr(C2O4)4−4). Initial experiments to determine the Kd of 88Zr tracer onto the magnetic nanoparticles from oxalic acid resulted in low Kd's (on the order of ˜100 mL/g). This distribution coefficient is insufficient for effective isotope loading. Next, the Kd's of 88Zr oxalate on the magnetic nanoparticles were determined after the solution was buffered using physiological saline and HEPES buffer. An increase in 88Zr tracer binding was observed, with Kd values up to 4700 and 46,000 mL/g for FeO and MnFeO, respectively. This represented a substantial improvement in uptake. However, the Kd values were still substantially lower than those observed in the ZrCl4 system. Therefore, the team evaluated a method to convert the 89Zr oxalate product to the chloride form.
In order to obtain the very high Kd values for Zr on the magnetic nanoparticles, as shown in
A modified method was devised to elute Zr from the AnIX column using HCl concentrations as low as 0.15 M HCl. This was accomplished by inserting an intermediate step in the AnIX separation process. Zr(C2O4)4−4 was loaded onto a micro-column (27 μL column) packed with a strongly basic anion exchange resin (QAE-550C) that had been previously converted to the oxalate form). Following the load, the column was washed with clean 0.8 M H2C2O4, and then the column was converted to the formate form via the addition of formic acid. Zirconium has a high level of retention on the AnIX column in formate form, so it remains fixed on the column during this step.
The formate anion (HCO2−) has a low “relative selectivity coefficient” on the AnIX resin. For MP-1M resin, formate has a selectivity coefficient of 4.6. This compares to a value of 22 for the chloride anion. An oxalate value is not specified directly in the literature, but it is estimated to be very high—likely somewhere below citrate (220) and above iodide (175). By converting the column from oxalate form (˜200) to formate form (4.6), the Zr can be eluted from the AnIX column easily using moderate concentrations of Cl− ion.
Given the new column method to produce ZrCl4 from Zr(C2O4)4−4, efficient labeling of the magnetic nanoparticles was now possible. The above chemical method was used to convert purified 88Zr tracer from the oxalate form to the chloride form as a necessary step in order to successfully label the isotope onto the magnetic nanoparticles. 88Zr tracer was eluted with HCl concentrations of 0.15 M and 0.3 M. Additionally, the 88Zr tracer was eluted with 0.15 M and 0.3 M HCl that was diluted 1:1 with phosphate buffered saline (PBS). Prior to conducting batch contact experiments to determine the distribution coefficients across the various matrices, each of the four 88Zr tracer eluent solutions was diluted again into PBS. The final PBS dilution, resulting pH, and chloride concentration is indicated in Table 14.
Measured distribution coefficient values are presented as Log Kd; percent 88Zr tracer uptake values are listed in parentheses.
Within the solutions, which ranged between pH 1 and pH 2, the FeO and MnFeO nanoparticles were observed to be relatively stable over the contact period. For MnFeO MNPs, a 1 h contact time resulted in <1% Fe mass solubilization and ˜17% Mn mass solubilization fractions. Despite the loss of Mn from the MnFeO MNP media, however, the material was still capable of analyte sorption.
Distribution coefficient values for both nanoparticle types were very high, with Kd's typically measured at levels exceeding 10,000 mL/g for FeO and exceeding 100,000 mL/g for MnFeO. The Kd values are in good agreement with those observed in a matrix of pH-controlled physiological saline (
The superparamagnetic metal oxides Fe3O4 and Mn-doped Fe3O4 were shown to exhibit high affinities for Zr from dilute HCl solutions and solutions of physiological saline across a pH range. Measured affinities in oxalic acid were low, but were improved with the addition of buffering agents.
In accordance with example implementations, wherein 89Zr is eluted in an oxalate form, for example from a hydroxamate resin alone or from a hydroxamate resin as part of the dual column system described herein, it may be converted to a chloride form. A column method was developed in order to convert the 89Zr oxalate product fraction to 89ZrCl4, using formic acid to enable efficient 89Zr elution from the column using a low concentration of HCl. This resulting eluent solution, diluted into PBS, demonstrated that 89Zr Kd values were high for FeO (10,000 to >100,000 mL/g) and very high for MnFeO (>100,000 mL/g).
In compliance with the statute, embodiments of the invention have been described in language more or less specific as to structural and methodical features. It is to be understood, however, that the entire invention is not limited to the specific features and/or embodiments shown and/or described, since the disclosed embodiments comprise forms of putting the invention into effect. The invention is, therefore, claimed in any of its forms or modifications within the proper scope of the appended claims appropriately interpreted in accordance with the doctrine of equivalents.
Claims
1. A method for radio labeling monoclonal antibodies, the method comprising preparing a binding conjugate by exposing a ligand to a solution of 89Zr and less than 1 M oxalic acid.
2. The method of claim 1 wherein the solution comprises from 0.3 M to less than 1 M oxalic acid.
3. The method of claim 1 wherein the solution further comprises fluoride.
4. The method of claim 3 wherein the solution further comprises NaF.
5. The method of claim 1 wherein the conjugate comprises mAbs.
6. The method of claim 5 wherein the conjugate comprises deferoxamine.
Type: Application
Filed: Aug 30, 2021
Publication Date: Dec 16, 2021
Applicant: Battelle Memorial Institute (Richland, WA)
Inventor: Matthew J. O'Hara (Richland, WA)
Application Number: 17/460,986